Accuracy class
0.04
0.015
0,008
% Emax
Maximum capacity
Highest limit of specified measuring range
Emax
See Ordering information table
See Ordering information table
See Ordering information table
kg
Safe load limit
Maximum load without irreversible damage
Elim
For 500 kg … 50 t: 150
For 60 t: 125
For 75 t: 100
For 500 kg … 50 t: 150
For 60 t: 125
For 75 t: 100
For 500 kg … 50 t: 150
For 60 t: 125
For 75 t: 100
% Emax
Destructive load
Danger of mechanical destruction
Ed
For 500 kg … 50 t: >300
For 60 t: >250
For 75 t: >200
For 500 kg … 50 t: >300
For 60 t: >250
For 75 t: >200
For 500 kg … 50 t: >300
For 60 t: >250
For 75 t: >200
% Emax
Minimum LC verification
Minimum load cell verification interval (vmin = Emax/Y)
See Accuracy classes and minimum verification interval, vmin table
See Accuracy classes and minimum verification interval, vmin table
See Accuracy classes and minimum verification interval, vmin table
For Emax = 500 kg
Y
2,500
n.a.
n.a.
For Emax = 1 t
Y
5,000
n.a.
n.a.
For Emax = 2 t
Y
5,000
10,000
n.a
For Emax = 3 t … 75 t
Y
5,000
14,000
3 t – 10 t: 14,000
>20 t: 20,000
Rated output
Relative output at maximum capacity
Cn
2
mV/V